21 CFR Part 211 · Current Good Manufacturing Practice for Finished Pharmaceuticals

§ 211.160 General requirements

Subpart I - Laboratory Controls Official text on eCFR Amendment history

(a) The establishment of any specifications, standards, sampling plans, test procedures, or other laboratory control mechanisms required by this subpart, including any change in such specifications, standards, sampling plans, test procedures, or other laboratory control mechanisms, shall be drafted by the appropriate organizational unit and reviewed and approved by the quality control unit. The requirements in this subpart shall be followed and shall be documented at the time of performance. Any deviation from the written specifications, standards, sampling plans, test procedures, or other laboratory control mechanisms shall be recorded and justified.

(b) Laboratory controls shall include the establishment of scientifically sound and appropriate specifications, standards, sampling plans, and test procedures designed to assure that components, drug product containers, closures, in-process materials, labeling, and drug products conform to appropriate standards of identity, strength, quality, and purity. Laboratory controls shall include:

(1) Determination of conformity to applicable written specifications for the acceptance of each lot within each shipment of components, drug product containers, closures, and labeling used in the manufacture, processing, packing, or holding of drug products. The specifications shall include a description of the sampling and testing procedures used. Samples shall be representative and adequately identified. Such procedures shall also require appropriate retesting of any component, drug product container, or closure that is subject to deterioration.

(2) Determination of conformance to written specifications and a description of sampling and testing procedures for in-process materials. Such samples shall be representative and properly identified.

(3) Determination of conformance to written descriptions of sampling procedures and appropriate specifications for drug products. Such samples shall be representative and properly identified.

(4) The calibration of instruments, apparatus, gauges, and recording devices at suitable intervals in accordance with an established written program containing specific directions, schedules, limits for accuracy and precision, and provisions for remedial action in the event accuracy and/or precision limits are not met. Instruments, apparatus, gauges, and recording devices not meeting established specifications shall not be used.

Independent copy, not an official legal edition. 21 CFR is a U.S. Government work in the public domain; this text comes from the eCFR and the annual CFR edition, and may lag recent amendments. The authoritative version is on ecfr.gov. Compliome is not affiliated with or endorsed by FDA. Not legal advice.

What FDA inspectors cite under § 211.160

Form FDA 483 observations citing this section, fiscal years 2009–2026 (Drugs).

  • Scientifically sound laboratory controls 1,444
  • Following/documenting laboratory controls 358
  • Calibration - at intervals, written program, remedial action 292
  • Lab controls established, including changes 218
  • Establishment of calibration procedures 143
  • Deviations from laboratory control requirements 134

Warning letters citing § 211.160

193 FDA warning letters since 2017 cite this section ; the most recent:

FDA guidance that cites § 211.160

and 23 more.

Amendment history

Federal Register documents that created or amended this section.